Menu
NEONATAL JAUNDICE PHOTOTHERAPY DEVICE

NEONATAL JAUNDICE PHOTOTHERAPY DEVICE

Features:

1. LED tube as jaundice irradiation components.

2. The Angle of the irradiation light box is adjustable.

3. The height of the irradiation light box is adjustable.

4. Equipped with a timer.

5. Castor is lockable

6. Stainless steel column.

7. Use with infant incubator.


Technical Parameters:

General

 parameters

Model

BHZ-001(Mobile)

BHZ-002

(Incubator Mounted)

Power Supply

AC 220V,50Hz, 75VA  

Light Source

LED

Spectral range

425nm~475nm

Lifetime of the light source

5000 hours

Treatment distance

40cm

46cm

The highest total irradiance for bilirubin Ebi max on 

the effective surface area

4.0mW/cm2±10%

2.6mW/cm2±10%

The total spectral irradiance on the effective surface area

≥1.6mW/cm2

≥1.0mW/cm2

Total irradiance for bilirubin between all points Ebi 

min Ebi max measured on the effective surface area

≥2.5mW/cm2

≥1.3mW/cm2

Uniformity G2of the total irradiance for bilirubin

>0.4

Infrared radiation on the effective surface area

≤1mW/cm2

Ultraviolet radiation on the effective surface area

≤1.0*10-5mW/c2(180nm<λ≤400nm)

Working noise

≤45dB(A)

Adjusting range for the radiation angle of lamp module

0-60°up from horizontal

/

Adjusting range for the height of lamp module

1350mm~1650mm

/

Built-in timer range

0~9999 hour 59 minute

Environment

 Temp

Operating range

28℃~34℃

Storage and transport range

-40℃~+55℃

Environment Humidity

Operating range

30%~75%

Storage and transport range

≤93%

 

Atmosphere

 Pressure

Storage and transport atmospheric pressure range

500hPa~1060 hPa

Working atmospheric pressure range

800 hPa~1060 hPa

Application environment altitude

≤2000m

Overvoltage category

Pollution degree

2

Air flow rate

Ambient air movement rate

<0.3m/s

Other Specification

Phototherapy Unit Measurement (L*W*H mm)

705*620*1350

1000*420*810

Phototherapy Unit Weight (kg)

18

10

The lifetime

6 years


The total spectral irradiance Ebi over a wavelength interval of 5nm for the wavelength range between 320nm and 550nm.

320

0.0000000000

380

0.0000000000

440

2.0851640000

500

1.5803700000

325

0.0000000000

385

0.0000000000

445

4.0588730000

505

0.9958963000

330

0.0000000000

390

0.0000000000

450

7.1860800000

510

0.6231434000

335

0.0000000000

395

0.0000000000

455

11.5261200000

515

0.3886996000

340

0.0000000000

400

0.0000000000

460

18.2637400000

520

0.2505039000

345

0.0000000000

405

0.0000000000

465

21.0351500000

525

0.1611249000

350

0.0000000000

410

0.0000000000

470

19.3089700000

530

0.1039433000

355

0.0000000000

415

0.0014880750

475

13.6934800000

535

0.0651738000

360

0.0000000000

420

0.0412178700

480

9.3698260000

540

0.0426970300

365

0.0000000000

425

0.1388389000

485

6.3107570000

545

0.0260784500

370

0.0000000000

430

0.3533474000

490

3.9971590000

550

0.0148105000

375

0.0000000000

435

0.8143584000

495

2.4550760000


 


Spectral distribution


Testing device integral Ebi calibration trend