Menu
INFANT INCUBATOR BK-IN 2/3/4

INFANT INCUBATOR BK-IN 2/3/4

The infant incubator provides newborns with a temperature and humidity adjustable environment similar to that of the mother's womb, which does not apply to the transfer of the baby. 

1.Microprocessor based servo control mode: air temperature, skin temperature, humidity, oxygen concentration.
2.Humidity is adjustable in two levels.
3.LED display: set temperature, air temperature, skin temperature, humidity, heat power rate.
4.Self–testing function;various audible and visual alarms.
5.>37℃ preset mode.
6.Independent locking device for front door.
7.The inclination of infant bed is adjustable.


Model

BK-IN 2

BK-IN 3

BK-IN 4

Specification of hood

Hood size (W*D*H)mm

980*630*740

Mattress size(W*D)mm

650*370

Thermostat Cover

Single-layer

Double-layer

Access Port

6(2 Iris Ports)

Operating Window

4

Infusion Seal

3

5

Technical Parameters

Power Supply

AC110/220V±10%, 50/60Hz

Consumption

650W

Display

LED

Control Model

Air

Air; Skin

Air Temp. Control Range(℃)

25~37, 37.1~38

Skin Temp. Control Range(℃)

N/A

32~37, 37.1~38

Skin Sensor Temp. Accuracy(℃)

N/A

±0.3

Temp. Rise Time(from 25 ℃)

<30min

Temp. Fluctuation(℃)

±0.5

Mattress (Horizontal) Temp.  Uniformity (℃)

≤0.8

Relative Humidity Display Range

N/A

0~100%RH

Relative Humidity Control Range

N/A

40%~95%RH

Oxygen Contentration Diplay Range

N/A

0~100%

Noise Level of Inside Incubator

<50dB[A]

Infant Bed Tilt Angle

±10%

Operating Condition

Working Temp. Range(℃)

20~30

Working Relative humidity

30%~75%RH

Ambient Air Flow Rate

<0.3m/s

Working Atmospheric Pressure

500~1060hpa

Alarm

Power Failure, Fan Failure, Temp. Sensor Failure, Over AirTemp.Alarm,

Temp. Deviation

Standard Accessory

RS232 interface connector, IV pole, tray, drawer*4

Net Weight

80kg

95kg

100kg

Package Size(W*D*H)mm

Contro Unit and Cabinet

Hood

1260*710*850

(Control unit and base)

1120*660*980

  980*630*740

1120*660*980

  980*630*740

Gross Weight

100kg

140kg

150kg