Menu
INFANT INCUBATOR BK-IN 1 PRO

INFANT INCUBATOR BK-IN 1 PRO

The infant incubator provides newborns with a temperature and humidity adjustable environment similar to that of the mother's womb,which does not apply to the transfer of the baby.

Features:

1.Microprocessor based servo control mode: air temperature, skin temperature, humidity, oxygen concentration.

2.LCD display: set temperature, skin temperature, humidity, oxygen concentration, heat power rate.

3.Double-layer thermostat cover; 4 operating windows and 2 iris ports.

4.Self–testing function;22 kinds of audible and visual alarms.

5.>37℃ preset mode.

6.Independent locking device for front door.

7.The inclination of infant bed is adjustable.



Model

BK-IN 1 Pro

Specification of Hood

Hood Size (W*D*H)mm

865*455*490

 Mattress Size(W*D)mm

770*380

Thermostat Cover

Double-layer

Access Port

6(2 Iris Ports)

Operating Window

4

Infusion Seal

8

Technical Parameters

Power Supply

AC110/220V±10%, 50/60Hz

Consumption

1320W

Heater Consumption

Heating chamber: 290W; Water box:170W

Display

LCD

Control Model

Air, Skin, Humidity, Oxygen concentration

Heating Indicator

0~100%, 10% increments

Air Temp. Control Range(℃)

25.0~37.0; 37.0~38.0

Skin Temp. Control Range(℃)

34.0~37.0; 37.0~38.0

Skin Sensor Temp. Display Range(℃)

5.0~65.0

Skin Sensor Temp. Accuracy(℃)

±0.3

Temp. Rise Time(Ambient Temp. 22℃±2℃)

≤50 min

Temp. Fluctuation(℃)

≤0.5

Mattress(Horizontal)Temp. Uniformity(℃)

≤0.8

Mattress(Tilted)Temp. Uniformity(℃)

≤1.0

The Difference between the Average Air Temp. and the Set Temp.(℃)

≤1.5

Water Box Capacity(ml)

1200

Relative Humidity Display Range

0%~99% RH

Relative Humidity Control Range

CAH%RH~90%RH

Humidity Display Accuracy

±10%

Oxygen Concentration Display Range

0%~99%

Oxygen Concentration Display Accuracy

±5%

Oxygen Sensor Lifetime

10000 hours under 100% oxygen concentration                                                       

Noise Level of Inside Incubator

≤55dB[A]

Infant Bed Tilt Angle

±10%

Operating Condition

Working Temp. Range(℃)

20~30

Transport and Storage Temp. Range(℃)

-20~55

Working Relative Humidity

30%~75% RH

Transport and Storage Relative Humidity

≤93%RH

Ambient Air Flow Rate

<0.3m/s

Working Atmospheric Pressure

800hPa~1060 hPa

Transport and Storage Atmospheric Pressure

500 hPa~1060 hPa

Application Environment Elevation

≤2000m

Overvoltage Category

Pollution Grade

2

General Alarm

Power Failure, Fan Failure, Temp.Deviation, Sensor failure,Over Temperature,Water Shortage, etc  

System Alarm

Communication Deviation, Temp. and Humidity Deviation

Standard Accessory

RS232 interface connector, X ray cassette, IV pole, tray, drawer*3.

External Size(W*D*H)mm

1050*675*1400~1600

Net Weight

117kg

PackageSize(W*D*H)mm

Control Unit and Hood

Base

 

1190*780*890

1100*760*830

Gross Weight

127kg