BK6400 Auto 5 parts Hematology analyzers are used to run tests on blood samples. They are used in the medical field to do white blood cell counts, complete blood counts, reticulocyte analysis.
1. 28 parameters, 2 histograms, 2 scattergrams
2. 60 tests/hour
3. 2 test modes & 2 sample modes
4. Integral titanium incubation system
5. 3-D topographic maps for WBC
6. Constant linear sheath flow
7. Laser Scatter Technology
Model | BK-6400 | |||
Throughput | 60 tests/hour | |||
Assay Items | 5 parts, 28 parameters, 2 histograms, 2 scattergrams | |||
Test mode | CBC+5DIFF, CBC | |||
Measuring Principle | Laser Scatter Technology | |||
Testing Items | Regular Parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, PDW, MPV, PCT, P-LCR, NEU#, NEU%, LYM#, LYM%, MON#, MON%, EOS#, EOS%, BAS#, BAS% | |||
Research Parameters: ALY#, ALY%, IG#, IG% | ||||
Performance | Item | Linear range | Carry over | CV |
WBC | 1.0-99.9x10^9/L | ≤0.5% | ≤2.0% | |
RBC | 0.3-7.0x10^12/L | ≤0.5% | ≤1.5% | |
HGB | 20-240g/L | ≤0.5% | ≤1.5% | |
PLT | 20-999x10^9/L | ≤1.0% | ≤4.0% | |
Storage | 100,000 results storage including histogram and scattergram | |||
LIS Interface | LIS Interface with HL7 protocol | |||
QC and Calibration | Multiple QC rules, including L-J, X-B etc. Auto and manual calibration function for both whole blood mode and pre-diluted blood mode | |||
Incubation | Integral titanium incubation system | |||
Operation Environment | Temperature: 15℃-30℃ | |||
Humidity: 30%-85% | ||||
Power supply | 110/220V±10% , 60/50Hz | |||
Package Size | 750*650*910mm | |||
Gross Weight | 80kg |