Menu
BK-240(PEARL) AUTO CHEMISTRY ANALYZER

BK-240(PEARL) AUTO CHEMISTRY ANALYZER

①.40 Sample positions 


②.80  Reagent positions  


③.44 Reaction cuvettes  

  

④.200 Tests/ Hour 



Sample/Reagent tray
·40 sample position
·80 reagent position
·Reagent refrigerated function


Reaction tray

·44 reusable cuvettes

·Automatic blank testing for cuvettes

·Colorimetry system of the pattern of light splitting behind cuvettes

·Dismountable, easily replacement

·Reaction temperature 37,constant temperature function



Sample/Reagent Probe

·Automatic liquid level detecting

·Collision protection function

·Probe internal/external washing

·Probe depth automatic adjustment



Auto cleaning system

·Automatically 4-step wash

·3 times repeat wash

·Hi-tech washing station to ensure

accurate results



Auto mixer, independently

·Automatically mix reagent and sample in cuvettes

·Ensure full reaction

·Automated wash itself

·Avoid cross-contamination



Perfect standard diagram

·Automatic show reaction curve

·QC result diagram recorder,print available



Intelligent software

·Automatic washing cuvette 

when start test&finish test

·Automatic detect sample& reagent volume

·Reaction temperature detecting when startup

·Alarm: automatic alarm of no reagent,no sample,wrong run,out of linearity range etc

·Automatic re-test sample in next order after no reagent



Real-time and Dynamic display running status

·Real-time show the running status

·Real-time show the status of reagent tray,sample tray,cuvettes 

·Real-time update the reagent residual volume


Overall

Performance

Model

 BK-240(PEARL)

Type

 Vertical,Discrete,Fully automatic

Throughput

 200T/H

Analysis Method

 End-point,kinetic,Two-point

Certification

 CE, FDA, ISO9001, ISO14001, ISO13485

Sample & Reagent Unit 

Sample Positions

 40 sample positions, including calibration, quality control and emergency

Reagent Positions

 80 refrigerated reagent positions

Sample Volume

 1ul ~30ul (0.1ul stepping)

Reagent Volume

 R1&R2:20ul ~300ul (1ul stepping)

Sample& Reagent Probe 

Automatic liquid level detecting,Collision protection function,Probe internal/external washing,Probe deepth automatic adjustment

Probe cleaning

 Automatic washing both interior and exterior

Reagent tray cooling

 Continuous 2℃~8℃ cooling within 24 hours

Reaction Unit

Reaction Time

 10 minutes

Temperature Control

 37℃±0.1℃, with online software monitor the temperature

Cuvettes

 44 reusable reaction cuvettes (optical path length 6mm)

Washing

 After finishtesting, washing probe wash cuvettes automatically

Optical System

Light Source

Halogen-tungsten lamp

Wavelength

300-700nm,8 wavelengths, Precision±1nm (4 positions are optional)

Absorbance range 

0-4.00Abs

Linear Range

0.0001Abs

Calibration and QC

Calibration

 Linear/nonlinear,one point,multi-point(Max 8 points)

Quality Control

QC plot automatically depict the function condition,show the QC results

Software

Intelligent software

Automatic washing cuvette when start test&finish test,Automatic detect sample& reagent volume,Reaction temperature detecting when startup

Alarm

Automatic alarm of no reagent,no sample,wrong run,out of linearity range etc, Automatic re-test sample in next order after no reagent

Software

 Windows 7,Windows 8 or above, windows XP

Others

Power Supply

 ~100-240V, 50-60Hz,1000VA

Package Size(W*D*H)

 870*720*1320mm

Gross Weight

 136kg