GAS CHROMATOGRAPH 1

GAS CHROMATOGRAPH 1

Features:

1. 7-inch LCD display indicates temperature, gas flow and heating curve.

2. Self-test function and automatic fault recognition function.

3. 8 pcs external extension interfaces can connect with control valve.

4. Equipped with a flame ionization detector (FID) which adopts APC control.

5. Multi-core and 32-bit embedded hardware system ensures the reliable operation.

6. Synchronous external trigger function and the external signal (auto sampler, hot analytical instruments, etc.) can start the equipment and workstation software at the same time.

7. With chromatography workstation software to process data. 


Model

BK-GC112A

Oven

Temp. Range

RT+5℃~400℃

Temp. Accuracy

±0.02℃

Temp. Program

9-phase/ 10 platform

Max Temp. Heating Rate

60℃/min

Sample injector

Temp. Range

RT+7℃~420℃

Column or split injector can be installed.

Split mode can directly set the split ratio.

Constant pressure mode.

A separate injection system is connected.

Detector

Temp. Range

RT+7℃~420℃

Hydrogen Flame Ionization Detector (FID)

Logarithmic amplifier

High voltage switch control

Baseline signal display

Ignition coil control

Detection limit:≤3*10-12g/s

(Sample: n-hexadecane)

Dynamic range: 107

Thermal Conductivity Detector (TCD)

(Optional)

Bridge voltage switch control

Bridge current setting: 0-220mA

Sensitivity: 6000mv.ml/mg

Dynamic range: 105

Power Supply

AC220V±10%, 50/60Hz

Standard Accessory

FID detector, Capillary injection system, Chromatography workstation software.

Optional Accessory

TCD detector, Chromatographic column, Gas purifier.

Package Size (W*D*H)

1020*660*740mm

Gross Weight

82kg