Menu
CLOSED-CUP FLASH POINT TESTER

CLOSED-CUP FLASH POINT TESTER

It is used to make determination of the closed cup flash point of the petroleum products.

Features

BK-FP261

* The heating power is adjustable and shown visually by voltmeter.

* The structure is designed small and exquisite with stainless-steel table board

BK-FP261A

* Simulation tracking displays function curves of heating up and test time.

* Automatically calibrate the effect of atmospheric pressure to test the corrected value.

* Differential coefficient detection.Automatically correct the system deviation.

* Automatically open the lid,detect the flash point and print test data.

* The test arm rises up and lowers down automatically.

* Igniting with electric igniting gun,forced air cooling.



Model

BK-FP261

BK-FP261A  

Ambient Temperature

≤ 35℃

≤ 40℃

Relative Humidity

≤85%

≤80%

Consumption

650W

500W

 

Flash Point Determination

Mercury-in-glass thermometer.Specifications are as below:

(1)Scale -5℃~110℃, division 0.5℃

(2)Scale 20℃~150℃, division 1℃

(3)Scale 90℃~370℃, division 2

Range: Ambient to 300℃

Repeatability: ≤3℃

Reproducibility: ≤4℃

Accuracy: 0.1℃

 

 

Oil Cup

- Inner diameter: 50.7mm~50.8mm

-Depth: 55.7mm~56.0mm

- The scribed line depth of capacity of testing oil: 33.9mm~34.3mm

-Capacity of testing oil: about 70ml

Stirring Rate

Procedure A: (90~120)RPM, Procedure B: (250±10)RPM

Igniting Device

 

(1) Igniting source: gas(or other civilian fuels, the same below)

(2) Electric ignition. Gas flame diameter: 3.2mm~4.8mm

Heating Device

The heating power is continuously adjustable from 0 W to 600W

Heating Rate

(1~12)℃/min; Controllable and adjustable

Procedure A: (5~6)℃/min,

Procedure B: (1~1.6)℃/min

Power Supply

AC220V±10%, 50Hz

Dimension(W*D*H)

340*320*450mm

520*360*310mm

Packing Size (W*D*H)

470*400*410mm

600*510*600mm

Gross Weight

12.5kg

25kg